• ISO-17025 Accreditation

    Maintenance & Calibration

generic-media-pages-sm-17191-maintenance-and-calibration-iso17025-accreditation-0-0.webp


ISO/IEC 17025:2017


ISO/IEC 17025 describes the general requirements for the competence of testing and calibration laboratories. The aim of this standard is to increase confidence in the work of laboratories by describing precise requirements to deliver competent work and valid, internationally comparable results. The standard also includes requirements to address risks and opportunities in order to increase the effectiveness of the management system and thus lead to improved results.

The European Konica Minolta Service Center in Bremen, Germany, is accredited to the latest version (ISO/IEC 17025:2017) of this standard. As Konica Minolta is a global company, A2LA (American Association for Laboratory Accreditation) has been chosen for accreditation in Europe, as well as the USA and Japan.

In short, this accreditation provides you with the assurance that you can rely on our services and obtain comparable measurement results worldwide.

Scope of accredited calibrations

Parameter/Equipment

Range

Comments / Instruments

Illuminance – Illuminance Meters

(8, 80, 800) lux

T-10A

T-10MA

CL-200A

CL-500A

Luminance – Luminance Meters

(15, 50, 100, 150, 200, 230) cd/m2

LS-150

LS-160

Correlated Color Temperature (CCT) – Illuminance Meters

2856 K

CL-200A

CL-500A

CS-150/160

Chromaticity (x, y) – Illuminance Meters

0.4457 £ x £ 0.4488

0.4069 £ y £ 0.4078

CL-200A

CL-500A

Chromaticity (x, y) – Luminance Meters

0.4457 £ x £ 0.4488

0.4069 £ y £ 0.4078

CS-150

CS-160

di:8° (SCI) Spectral Reflectance Factor

(80 to 100) %

360 nm

(370 to 390) nm

(400 to 470) nm

(480 to 720) nm

730 nm

740 nm

Reflectance standard tile

CM-3700A

CM-36d(GV)

CM-5

CM-26d(G)
CM-25d

CM-23d

CM-700/CM-600

de:8° (SCE) Spectral Reflectance Factor

(80 to 100) %

360 nm

(370 to 390) nm

(400 to 470) nm

(480 to 720) nm

730 nm

740 nm

Reflectance standard tile

CM-3700A

CM-36d(GV)

CM-5

CM-26d(G)
CM-25d

CM-23d

CM-700/CM-600

45°:0° Spectral Reflectance Factor

(80 to 100) %

(360 to 390) nm

(400 to 450) nm

(460 to 720) nm

730 nm

740 nm

Reflectance standard tile

CM-25cG

Link to certificate:

TECHNICAL SERVICE